A 72-year-old patient with a diagnosis of nonfluent progressive aphasia, who lived independently and was felt to have good judgment, received a biventricular implantable cardioverter-defibrillator (ICD) in the setting of unexplained syncope and cardiomyopathy. He presented to the emergency department 2 months later with complaints of recurrent ICD discharges. His device interrogation revealed more than 60 ICD shocks over the previous week, with rapid irregular electrical activity on the right ventricular (RV) lead, suggestive of atrial fibrillation sensing. The RV lead was not sensing ventricular activity or capturing the ventricle. The ICD pacing and therapies were turned off, and the patient was admitted for further investigation. Chest x-ray (Fig. 1) revealed retraction of the RV lead to the midatrial location and retraction of the atrial lead to the subcutaneous pocket. The leads were also notably entangled and twisted around each other near the device. When confronted with these findings, the patient admitted to repeatedly manipulating the device. He was taken for a lead revision procedure, during which a photograph was obtained (Fig. 2), which shows the appearance of the entangled leads caused by the patient's repeated twisting and turning of the device.

Fig. 1Fig. 1Fig. 1
Fig. 1 Chest x-ray obtained on presentation (solid arrow = right ventricular lead in the right atrium; open arrow = right atrial lead).

Citation: Texas Heart Institute Journal 49, 6; 10.14503/THIJ-21-7702

Fig. 2Fig. 2Fig. 2
Fig. 2 Entangled leads postexplanation.

Citation: Texas Heart Institute Journal 49, 6; 10.14503/THIJ-21-7702

Comment

Twiddler syndrome refers to a patient's recurrent manipulation of the pulse generator, either consciously or subconsciously, which often leads to lead dislodgement and device malfunction.1,2 This phenomenon was described soon after patients started receiving pacemakers.3 The loss of contact of the leads with the cardiac tissue resulting from retraction can result in a fatal loss of pacing in pacemaker-dependent patients or recurrent inappropriate device therapies, such as those described in the presenting patient. An excessively mobile device, obesity, pocket discomfort, and psychiatric/cognitive disorders are factors associated with this clinical entity. There are several preventive measures to be considered: careful anchoring of the device and its leads, submuscular device implantation, nonabsorbable antimicrobial device envelopes, and ensuring appropriate pocket size.4

Twiddler syndrome should be on the differential whenever lead malfunction is suspected, especially soon after a device implant.

Conflict of Interest Disclosures: All authors have no conflicts to disclose.

Funding/Support: This research did not receive any specific grant from funding agencies in the public, commercial, or notfor-profit sectors.

References

  • 1.

    Weir RAP, Murphy CA, O'Rourke B, Petrie CJ. Twiddler's syndrome: a rare cause of implantable cardioverter defibrillator malfunction. Eur Heart J.2016;37(

    46
    ):3439. doi:10.1093/eurheartj/ehv383

  • 2.

    Chaara J, Sunthorn H. Twiddler syndrome. J Cardiovasc Electrophysiol.2014;25(

    6
    ):659. doi:10.1111/jce.12337

  • 3.

    Bayliss CE, Beanlands DS, Baird RJ. The pacemaker-Twiddler's syndrome: a new complication of implantable transvenous pacemakers. Can Med Assoc J.1968;99(

    8
    ):371373.

  • 4.

    Osoro M, Lorson W, Hirsh JB, Mahlow WJ. Use of an antimicrobial pouch/envelope in the treatment of Twiddler's syndrome. Pacing Clin Electrophysiol.2018;41(

    2
    ):136142. doi:10.1111/pace.13259

  • Download PDF
Copyright: © 2022 by the Texas Heart Institute, Houston
Fig. 1
Fig. 1

Chest x-ray obtained on presentation (solid arrow = right ventricular lead in the right atrium; open arrow = right atrial lead).


Fig. 2
Fig. 2

Entangled leads postexplanation.


Contributor Notes

Corresponding author: Seri Hanayneh, MD, Section of Cardiology, Tulane School of Medicine, 1430 Tulane Avenue, New Orleans, LA 70112 E-mail:shanayneh@tulane.edu